The Circulator features and advanced PID temperature controller which can be used in either following ways:
Suggested applications:
The Thermocirculator is supplied assembled to a stainless steel mounting bridge which features 'push-fit and lock fittings' for secure external liquid circulation. The bath consists of a 304 spec stainless steel tank and heat exchange coil for optimal cooling with a stainless steel shelf all giving good temperature stability.
Supplied with flat lift-off lid with insulated handle which can be substituted with accessory SL4 Gable Lid for immersion of taller samples in the bath.
Construction
Performance information, as the temperature decreases the cooling power of the ChilloBath reduces. ChilloBath uses following refrigerant R134A.
Guidance on Circulator
Performance of flow and pressure is affected by restrictions: type circulating liquids, height, pipe length and bore, temperature of the liquid falls below 5°C the flow reduces and cooling rate falls.
Guidance on Chiller Circulators and a Chiller recirculators.
A chiller circulator is suitable when temperature stability is the desired result, a chiller recirculator is used for heat removal.
Guide Specific Applications.
Guidance on additional cooling Can use accessory DC1-300 to boost the cooling capacity of the water bath.
Product Name : |
NE7-8 |
|||||||||||||||||||||||
Product Description : |
8 Litre ChilloBath |
|||||||||||||||||||||||
Nominal Tank Capacity : |
6 Litres |
|||||||||||||||||||||||
Size selector : |
||||||||||||||||||||||||
Product Specification : |
This Chiller Circulator range of water baths are specially designed for circulating temperature controlled liquids maintaining temperature in analytical equipment, heat removal or immersion of samples. The ChilloBath consists of a stainless steel tank fitted above a cooling system with a Thermocirculator immersed into tank. All immersed components are corrosion resistant. The Thermocirculator features: |
|||||||||||||||||||||||
Brimful Capacity : |
8 Litres |
|||||||||||||||||||||||
Tank, immersed components : |
Immersed components are corrosion resistant |
|||||||||||||||||||||||
Internal Dimensions : |
129w x 298d x 150h mm |
|||||||||||||||||||||||
External Dimensions : |
451w x 332d x 780h mm |
|||||||||||||||||||||||
Temperature Range : |
-20 to 130°C |
|||||||||||||||||||||||
Sensitivity/temperature stability @45°C : |
±0.1°C @0°C |
|||||||||||||||||||||||
Uniformity @45°C : |
±0.01°C @0°C |
|||||||||||||||||||||||
Temperature setting/control : |
digital |
|||||||||||||||||||||||
Temperature display resolution : |
Display of either set or measured temperature in increments of 0.1°C from 20°C - 99°C |
|||||||||||||||||||||||
Heat Up Times : |
|
|||||||||||||||||||||||
Heating/Cooling information : |
1. Cool down times are stated from ambient using 100% glycol or Lab Bath solution. 2.Heat up times are stated from ambient 20°C.3. It is recommended that to achieve optimum performance and reduce energy consumption the bath should be covered with supplied Lid. 4. All temperature measurements and performance claims have been carried out in ambient temperature 20°C and installed correctly with adequate ventilation. 5. Cool down and heat up times above are based on performance data recording change in bath liquid temperature.
|
|||||||||||||||||||||||
Warranty Period : |
3 Years |
|||||||||||||||||||||||
Safety features : |
1. In use a Safety cut-out Float switch monitors the water level in the bath and when it drops below a safe working amount the heater switches off. 2. Over temperature alarm operates +4°C with heater cut off, indication. 3. Under temperature alarm operates -4°C with indication.4. A Safety Experiment protector can be set just above desired working temperature protecting your work. 5. Wheels lockable when sited for use. 6. The Cooling system cuts out above 40°C. 7. Alround ventilation slots with lockable hinged access door to the condensor for easy cleaning. 8. Convenient push-fit and lock fittings - for secure external liquid circulation - reduces risk tubing becoming dislodged or leaking hot or cold fluids in use. |
|||||||||||||||||||||||
Safety Class : |
Class 2 - Low liquid cut out and Safety Experiment protector can be set just above desired working temperature protecting your work. |
|||||||||||||||||||||||
Heater Power (W) @ 220-240V : |
1250W |
|||||||||||||||||||||||
Power consumption (W) @ 220-240V: |
1300W and 600W |
|||||||||||||||||||||||
Voltage : |
230V |
|||||||||||||||||||||||
Drain tap fitted : |
Yes |
|||||||||||||||||||||||
Notes : |
Guidance on Cooling compressor capacity, ChilloBath |
|||||||||||||||||||||||
Features and Benefits : |
1. Easy to read bold temperature display to 0.1°C resolution (-20°C - 99°C). 2. Quick reliable selection and attainment of desired operating temperature. 3. Designed for good water circulation and water control stability in the tank. 3. Temperature alarms operates either +4°C or -4°C about set temperature. 4. A Safety Experiment protector can be set just above desired working temperature protecting your work. 6. Safety cut-out float switch - for low liquid level protection. 7. Fitted with wheels and handles allowing limited movement - wheels lockable when sited for use. 8. Quiet running powerful circulation pump operating at optimum speed for maximum circulation - closed loop. 9. Push, fit and lock tube circulation system which prevents the tube from becoming dislodged or leaking fluids - an important safety feature with hot and cold liquids. 10. A 304 spec stainless steel tank gives good resistance to many substances and a clean appearance throughout its working life. 11. Broad range of accessories available to suit wide range of applications. 12. Sturdy construction - powder coated exterior providing a durable chemical resistant. 13. ChilloBath has ventilation on all sides allowing it to be sited in any position for good air flow on all. 14. To maintain the full cooling performance a hinged lockable door allows access to the condensor to easily remove dust and dirt accumulated during normal use. 15. Built to latest safety standards and supplied with comprehensive operating instructions and Three-year warranty. 16. Each bath is uniquely identified with its own serial number recorded on our product database, which includes all manufacturing inspection data. Electrically tested for safety at final inspection and a test certificate enclosed in rear of the instruction book. 17. Each ChilloBath is tested to lowest temperature extreem to verify its performance of the temperature control system. 18. The Chiller Circulator undergoes 2 point calibration routine at -20°C and 65°C before it leaves the factory for operating assurance. 19. Carries the CE mark meeting the requirements of Low Voltage and EMC Directives. 20. Powder painted body providing a durable finish and protection against corrosion and now featuring “active anti-bacterial technology”. This paint finish actively inhibits bacterial growth improving hygiene requirements as demanded by many clinical and laboratory environments. Hygienic coatings are part of a controlled approach to cleaner facilities. The Clifton range has always featured easy to wipe clean smooth paint finishes that are resistant to routine and efficient disinfection’s, however this improved coating featuring ‘active anti-bacterial’ properties provides an extra layer of defence. This new anti-bacterial paint finish is effective against major strains of bacterium including E-coli and Staphylococcus aureus, resistant strain. Within its formulation an active ingredient is a proven anti-bacterial agent bound into the coating during application and curing to the product. The efficacy of the paint finish applied to the Clifton range is maintained over time, as the anti-bacterial agent is integral within the paint finish.
|
|
|